Why Is the "Calibrate Stage" Option Greyed Out Even Though My X-Y Stage Is Recognized?

Why Is the "Calibrate Stage" Option Greyed Out Even Though My X-Y Stage Is Recognized?

Why Is the "Calibrate Stage" Option Greyed Out Even Though My X-Y Stage Is Recognized?

The preferences for the X-Y stage are stored in two locations within Volocity:

  1. Device Preferences:

    • Accessible via Edit > Preferences > Devices (or Volocity > Preferences > Devices on macOS).
  2. X-Y Stage Preferences:

    • Accessible via Edit > Preferences > X-Y Stage (or Volocity > Preferences > X-Y Stage on macOS).

Troubleshooting Steps:

  1. Verify that the stage appears in the Device Preferences and that the X-Y Stage option is selected.
  2. If the "Calibrate Stage" function in the Stage menu is still greyed out, ensure that an appropriate X-Y stage device is selected in the X-Y Stage Preferences.

This ensures that Volocity recognizes the correct stage settings and enables the calibration functionality.

    • Related Articles

    • How to Change the Spatial Calibration of a Dataset in Volocity

      How to Change the Spatial Calibration of a Dataset in Volocity Proper spatial calibration is essential for accurate quantitative analysis in Volocity. Each objective lens on the acquisition system should be calibrated using a stage graticule, which ...
    • Why Are Focus Positions Not Stored When Creating XY Points?

      Why Are Focus Positions Not Stored When Creating XY Points? If the focus position is not being stored when creating or reviewing XY points, it is likely because a Focus Device has not been set in the X-Y Stage preferences. Without a Focus Device ...
    • Restoring a corrupt Library in Volocity

      Restoring a corrupt Library in Volocity A Volocity library is an essential database that combines image data and metadata. If a library becomes corrupt, it may be difficult or impossible to fully restore the data without a proper backup. For this ...
    • Connecting a Leica DMi8 Microscope to Volocity

      Connecting a Leica DMi8 Microscope to Volocity To connect a Leica DMi8 microscope to Volocity, you need to configure the microscope's Virtual COM Port (VCP). The Leica SDK, which is included with the DMi8 upon delivery, installs the necessary driver ...
    • How to Derive Calibration Properties (µm/pixel) for Quantitative Analysis

      How to Derive Calibration Properties (µm/pixel) for Quantitative Analysis If you are performing quantitative analysis on image data, it is most accurate to retrieve the calibration properties from the acquisition system or re-acquire the data after ...