Proper spatial calibration is essential for accurate quantitative analysis in Volocity. Each objective lens on the acquisition system should be calibrated using a stage graticule, which ensures that the measurements correspond to real-world dimensions. For detailed instructions on performing this calibration with Volocity Acquisition, refer to page 151 of the Volocity User Guide.
However, if the correct calibration values (microns per pixel) are already known, these can also be manually entered.
Select the Image Sequence:
Access Image Properties:
Modify Calibration Values:
If the spatial calibration has not been performed using a stage graticule, you can calculate the µm/pixel value manually using the following equation:
µm/pixel = Camera Pixel Size (µm) / (Lens Magnification × Intermediate Magnification)
Scenario:
Calculation:µm/pixel = 6.45 / (63 × 0.63)µm/pixel = 0.163
Result:
The spatial calibration values for X and Y would be 0.163 µm/pixel, which can be entered into the µm/pixel (X) and µm/pixel (Y) fields in the Properties dialog.
While manual calculations can provide approximate calibration values, it is best practice to use a stage graticule for calibration. This ensures the calibration values are as accurate as possible, accounting for all variations in optical components and camera setup.