Why Can't I Change the Filter Turret Position with My Leica Microscope Using Volocity?
Why Can't I Change the Filter Turret Position with My Leica Microscope Using Volocity?
If the filter turret does not move to the transmitted light position when requested in Volocity, this is likely due to the Contrast Mode setting on Leica microscopes. Leica microscopes have contrast modes such as "Fluorescence," "DIC," or "DIC and Fluorescence," which can limit software control over the filter turret position.
Steps to Resolve the Issue:
Check and Adjust the Contrast Mode:
- Ensure that the Contrast Mode is set to the desired setting (e.g., "DIC and Fluorescence" for compatibility with both transmitted and fluorescent light paths).
Include the Contrast Mode in the Lightpath:
- Double-click on the desired lightpath button in the Video Preview.
- In the Device Checklist, ensure that Contrast Mode is checked.
- Click Save to update the lightpath settings.
If necessary, you can manually move the filter turret to the transmitted light position using the buttons on the microscope.
Related Articles
Connecting a Leica DMi8 Microscope to Volocity
Connecting a Leica DMi8 Microscope to Volocity To connect a Leica DMi8 microscope to Volocity, you need to configure the microscope's Virtual COM Port (VCP). The Leica SDK, which is included with the DMi8 upon delivery, installs the necessary driver ...
How to Change the Spatial Calibration of a Dataset in Volocity
How to Change the Spatial Calibration of a Dataset in Volocity Proper spatial calibration is essential for accurate quantitative analysis in Volocity. Each objective lens on the acquisition system should be calibrated using a stage graticule, which ...
Configuring the Serial Port of the DMi8 for Volocity
Configuring the Serial Port of the DMi8 for Volocity When you install the Leica SDK (delivered with the DMi8 microscope), it includes a driver that allows the computer to connect to the microscope via USB. This driver sets up a "Virtual Com Port" ...
Volocity Supported File Formats
Volocity Supported File Formats File Formats for Import Volocity supports the following file formats for reading: BioRad PIC Windows BMP Volocity Library Clipping DeltaVision ICS/IDS (Image Cytometry Format) JPEG Leica LIF (generated by Leica ...
Transferring Volocity Configurations Between Workstations or Users
Transferring Volocity Configurations Between Workstations or Users If you have purchased a new workstation or need to standardize settings across multiple user accounts on a single workstation, Volocity configurations can be transferred. Follow the ...